medRxiv preprint

Scan length as a major driver of CT radiation dose: a diagnostic reference level audit from Kosovo

Computed tomography is the largest contributor to population radiation dose from medical imaging, yet no diagnostic reference levels (DRLs) have been published from Kosovo or the Western Balkans. This retrospective audit analyzed all CT examinations performed on a 128- slice scanner at the University Clinical Centre of Kosovo between January and March 2026. After exclusions, 1,535 acquisitions from 1,092 patients across nine examination categories were analyzed. Local DRLs were defined as the 75th percentile and compared against German (BfS 2022) and Turkish (Kahraman et al., 2024) reference values. Head CT (n = 590) demonstrated CTDIvol 4.7% below the BfS DRL yet scan length 98.5% above the

radiology and imaging